The Capintec Model 815 kVp meter and exposure timer Measures the peak x-ray accelerating voltage from tungsten x-ray generators. To measure peak kV the user simply places the instrument in the beam and takes an x-ray. In addition to measuring kVp, the unit also measures the exposure time and displays the x-ray waveform type. The Capintec kVp Meter can be used on radiographic, fluoroscopic and dental x-rays. There is a high resolution mode which is used for measurements on older x-rays and x-rays with slow rise times. The kVp Meter has a large display which is readable from outside the x-ray room.
Features Small Hand Held Size Easy to use Non invasive High Accuracy Battery Operated Two Year Warranty Measures kVp and exposure time simultaneously Compact size Durable ABS Plastic Case Custom Carrying Case Option
Specifications Range 45 to 125 kV kVp Accuracy 2% +/- 1 kV, at 25 to 100 mA 18 to 42 cm from head, for Tungsten target x-ray tube with 1.5 mm Al equivalent filtration Exposure Time Accuracy 2% +/- 3 millisecond Minimum exposure time is 100 millisecond Minimum Exposure Time (kVp Measurement) 100 millisec - High Speed Mode (normal) 200 millisec - High Resolution Mode Minimum Current (mA) 7 mA at 50 kV, 10 cm from x-ray source Also depends on focus / collimation Display 0.4"(10.2mm) Liquid Crystal 8 Character Alphanumeric
Power
4 AA Batteries accessible from bottom of case Low Battery Indicator Battery Life 48 hours continuous Typically 9 months of normal use Controls/Indicators ON/OFF Switch Illuminated push-button with green power-on indication Mode Switch Momentary push-button 8 Character Liquid Crystal Display Oscilloscope output Physical Size 150 x 120 x 58.5 mm 5.9 x 4.7 x 2.3 inches Weight 0.7 kG (1.5 lb) Connections None required for normal kVp measurement 3.5mm Jack for Oscilloscope Output Optional Accessories Padded Cordura Carrying Case Oscilloscope Interface Cable
Typical Applications Measurement of Exposure Verification of peak x-ray voltage Constancy checks X-ray calibration Quality Assurance Measurement of X-ray Exposure Time Half Value Layer determination Troubleshooting and Repair of X-rays
|
|
|
|